LMS water

Dissolved Oxygen Measurement

LMS water offers the 501-OX Mackereth (galvanic) oxygen electrode, which includes a built-in Pt100 temperature sensor for automatic temperature compensation of the DO measurement. Each electrode is supplied with oxygen cell (OC3) "capsule" which is stored in a special canister.
The special storage canister for each oxygen cell (OC3) "capsule" performs three functions:
1. Provide protection from both physical damage and the ingress of atmospheric oxygen.
2. Short-circuit the capsule contacts: necessary for galvanic type oxygen cells.
3. A small sponge pad in the cap of the canister keeps the membrane moist. The sponge should be kept wet by periodically adding a few drops of distilled water.

The 501-OX oxygen electrode includes all the accessories for BOD (Biochemical Oxygen Demand) experiments.

Accessories

OC3: Replacement oxygen cell: 501-OX
SCD1: Sodium sulfite for 0% oxygen calibration
SED1: Electrolyte for 501-OX (galvanic)
{Dissolved Oxygen  Electrode}

Specifications

Dissolved Oxygen
Range/ Resolution 0.00 to 20.00ppm or mg/l, 0.0% to 200.0% Saturation
Accuracy (at T cal) ±0.2ppm or mg/l, ±2% Saturation
Temperature compensated
Accuracy ±0.3ppm, ±3% saturation (±10°C from T cal)
Temperature Range -8°C to 45°C
Compensation Keypad entry of salinity and barometric pressure